DCMPRTXXXIS

Data logger for temperature and pressure in hazardous areas
has been discontinued without replacement as of 01.08.2015.

Compact data logger for temperature and pressure
2400 baud

DatasheetTechnical supportRequest/On-site presentation
  • Can be used in potentially explosive atmospheres
  • Programmable start time
  • Real-time applications
  • Simple export of measurement data to EXCEL™
  • 1 year battery life
Datenlogger Datenlogger für Temperatur und Druck in explosionsgefährdeten Bereichen DCMPRTXXXXIS

The DCMPRTxxxIS data logger is a battery-powered data logger designed for use in potentially explosive atmospheres. The DCMPRTxxxxIS has FM certification and is classified in Class I, Division 1, groups A, B, C, and D and Class I, Division 2, groups A, B, C, and D. This certification allows the device to be used in sensitive environments such as air conditioning systems, cooling water, hot water, air, gas, oil and vapor pressure systems.
The integrated temperature sensor enables accurate temperature measurement without the need for an additional temperature data logger. The data logger has an NPT 1/4" pipe thread, which allows connection to most pressure lines.

Temperature sensor:

Integrated semiconductor sensor.

Temperature range:

-40 °C to +80 °C; resolution of 0.1 °C.

Accuracy:

±0,5 °C

Pressure sensor:

Semiconductor strain gages.

Print area:

0 PSIA/G to 30 PSIA/G (0 bar to 2.068 bar), 0 PSIA/G
to 100 PSIA/G (0 bar to 6.895 bar), 0 PSIA/G to 300 PSIA/G (0 bar to
20.684 bar), 0 PSIA/G to 500 PSIA/G(0 bar to 34.474 bar) , 0 PSIA/G
to 1000PSIA/G (0 bar to 68.948 bar), 0 PSIA/G to 5000 PSIA/G (0 bar
to 344,738 bar); depending on model.

Resolution:

0.002 PSIA/G (0.000138 bar), 0.005 PSIA/G (0.000345 bar), 0.02 PSIA/G (0.00138 bar), 0.05 PSIA/G (0.00345 bar)
bar), 0.02 PSIA/G(0.00138 bar), 0.05 PSIA/G (0.00345 bar); 0.05 PSIA
(0.00345 bar), 0.2 PSIA(0.0138 bar); depending on model.

Calibrated accuracy:

2 % of the measuring range, typically 0.25 % at 25 °C.

Pressure response time:

0.1 ms (10% to 90% of the measuring range).

Repetition rate:

±0.5 % of the measuring range; ±0.2 % of the measuring range
typical.

Storage:

16,383 measurements/channel; 32,766 measurements
total.

Recording rate

1 measurement/2 sec. to 1 measurement/12 hours.

Interface:

Serial or USB interface via external interface cable
interface cable (required as accessory), 2,400 baud.

Supply:

Replaceable 3.6 V lithium battery (included) with a typical
typical service life of 1 year at 25 °C.

Time accuracy:

±1 minute per month at 20 °C operating temperature,
without PC connection.

Data format:

Date; time and display of measured values in °C, °F, K,
°R, PSIA(g), inches, feet, mmHG, bar, Torr, kPa.

The data logger can only be used with the DCMIFC200 interface cable (USB)

Datenblatt
Data sheet DCMPRTxxxxIS

Software Datarecorder
Wachendorff DataRecorder

Project planning software for compact data loggers.

For device parameterization, data evaluation and data storage.

By downloading and using software and apps from this website, I agree to the GTC and the supplementary software clause.

Wachendorff Prozesstechnik Produktinformation
Product information card for DCMPRTxxxxIS